Healing Abutment Conical Connection RP 0 3.6 x 5mm, REF: 36644, UDI: 07332747034462 - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function. The Healing Abutment is intended to be used as a temporary component to an endosseous implant to allow healing of the soft tissue.

Company
Nobel Biocare Usa Llc
Recall Initiated
May 28, 2019
Terminated
February 9, 2021
Recall Number
Z-2250-2020
Quantity
234 pieces globally, of which 16 were sent to the US.
Firm Location
Yorba Linda, CA

Reason for Recall

Packaged dental implant healing abutments were found to contain the wrong size device.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of TN, IL, MI, and NY. The countries of AT, RU, BE, DE, GB, NO, PL, PT.

Lot / Code Info

Lot 13076845

Root Cause

Labeling mix-ups

Action Taken

On 5/28/2019, the firm mailed "Urgent Medical Device Recall" letters with an acknowledgement forms. The letter instructs the customers to: 1. Identify affected devices in your stock. 2. Do not use affected devices. 3. Complete and return the Customer Acknowledgment Form. 4. Return affected devices. If you require any further information or support, please contact your customer support representative at 1-800-322- 5001 x 1527.