RecallDepth

NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI (01)07332747091441(10)12129104(17)240203 - Product Usage: are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

Company
Nobel Biocare Usa Llc
Recall Initiated
August 28, 2019
Posted
May 4, 2020
Terminated
April 28, 2021
Recall Number
Z-1851-2020
Quantity
228 pieces
Firm Location
Yorba Linda, CA

Reason for Recall

The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).

Distribution

International distribution - the country Russia only.

Lot / Code Info

Lot:12129104

Root Cause

Labeling mix-ups

Action Taken

The firm's "Urgent Field Safety Notice", was issued on 08/28/2019 and the notices were sent to customers via email . The following actions are to be taken: 1) Identify affected devices in stock by confirming the information on the side label: 3.75x10mm, article no. 37965, lot 12129104. 2) Confirm that both the side and top label indicate 3.75x10mm. 3) Complete a Customer Acknowledgment Form and return it via email to: cs.russia@nobelbiocare.com within 5 days of receipt of the notice. 4) Return incorrectly labeled products (top label 3.75x7mm) to: Attn: Returns PFA1925 109004 Moscow Stanislavskogo str., h.21, b.3 (Correctly labeled products (top label 3.75x 10mm) do not need to be returned). 5) Returned devices will be replaced free of charge.

More recalls by Nobel Biocare Usa Llc

View all recalls by this company →