AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N

This recall has been terminated (originally issued August 27, 2019).

Company
Ad-Tech Medical Instrument Corporation
Recall Initiated
August 27, 2019
Terminated
April 5, 2023
Recall Number
Z-2544-2019
Quantity
315 units
Firm Location
Oak Creek, WI
Official Source
View on FDA website ↗

Reason for Recall

The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.

Distribution

NY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, PA, MD, and India, Korea, Taiwan, Canada, Hong Kong, Denmark, Russia, Singapore, Japan, Brazil. UK, Spain, Czech Republic, France

Lot / Code Info

All serial numbers distributed between January 2014 and May 2019

Root Cause

Under Investigation by firm

Action Taken

The firm initiated the recall by letter on 08/27/2019. The firm requested the units be returned.

More recalls by Ad-Tech Medical Instrument Corporation