AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode

This recall is currently active, issued October 24, 2024. It was issued by Ad-Tech Medical Instrument Corporation.

Company
Ad-Tech Medical Instrument Corporation
Recall Initiated
September 19, 2024
Posted
October 24, 2024
Recall Number
Z-0221-2025
Quantity
14 units
Firm Location
Oak Creek, WI
Official Source
View on FDA website ↗

Reason for Recall

Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.

Distribution

US Nationwide distribution in the states of CA, OH.

Lot / Code Info

UDI/DI 00841823107688, Lot Number 168788 855856857, 172999 873874875R, 174915 879880881

Root Cause

Employee error

Action Taken

AD-TECH issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 09/19/2024. The notice explained the problem with the devices, risk to health, and requested the consignee contact customer support to obtain an RMA to expedite the return of the affected product. For questions: contact your Ad-Tech Clinical Specialist.