Nucleic Acid test intended for screening for MRSA/SA. Each kit has 20 tubes. in vitro diagnostic.

Company
Roche Molecular Systems, Inc.
Recall Initiated
October 27, 2017
Terminated
April 27, 2020
Recall Number
Z-1195-2018
Quantity
136 kits
Firm Location
Branchburg, NJ

Reason for Recall

The cobas MRSA/SA tests may have decreased performance compared to the analytical sensitivity for MSRA detection. A screening failure can lead to potential infection in the colonized patient being tested as well as the spread of MSRA to others. The United States is not impacted, as the cobas MRSA/SA nucleic acid test for use on the cobas liat system is not available in the US.

Distribution

Worldwide distribution. Austria, Denmark, Germany, Singapore, Netherlands, Hong Kong, and Switzerland. US product is not affected.

Lot / Code Info

Lot #s: 70712A (6 kits); 70601W (76 kits); 70713A (4 kits); 70718A (50 kits); and 70809A (kits were not distributed to end users).

Root Cause

Software design

Action Taken

The cobas MRSA/SA tests has the potential for decreased performance compared to the analytical sensitivity for MSRA detection. A screening failure can lead to potential infection in the colonized patient being tested as well as the spread of MSRA to others.