RecallDepth

iTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101

Company
Conformis, Inc.
Recall Initiated
October 16, 2018
Terminated
June 29, 2020
Recall Number
Z-0731-2019
Quantity
1
Firm Location
Billerica, MA

Reason for Recall

The stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm.

Distribution

The products were distributed to the following US states: TN.

Lot / Code Info

Serial Number 0424647

Root Cause

Under Investigation by firm

Action Taken

The customer was contacted via email on 10/16/18.

More recalls by Conformis, Inc.

View all recalls by this company →