Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm, Material CA30L1

Company
Covidien Llc
Recall Initiated
August 7, 2017
Recall Number
Z-0560-2018
Firm Location
North Haven, CT

Reason for Recall

The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.

Distribution

Nationwide including PR, Canada, China

Lot / Code Info

UDI 10884521189553, Lot codes: S5LG004PX,S6AG001X,S6BG002X,S6BG005X,S6BG016PX,S6CG003X,S6CG005X,S6CG017X,S6DG004X,S6EG001X,S6FG001X,S6FG006X,S6FG010X,S6KG002X,S6KG006PRX,S6KG006PX,S6LG004X,S6MG002X,S7CG001PCX,S7CG011PCX,S7CG015CX,S7EG001X,S7EG003X,S7FG011PX

Root Cause

Device Design

Action Taken

All consignees were notified via Federal Express or certified mail on August 7, 2017, and the letter informs customers of the possible trocar tip disengagement and the actions they should take. Customers are requested to segregate and return affected product and acknowledge receipt and understanding of the Urgent Medical Device Recall Notice.