RecallDepth

PERI-LOC VOLAR DISTAL RADIUS - RIGHT 3H STD HEAD LOCKING PLATE, 62 MM LENGTH, SST PLATE, REF 71823114, QTY: (1) NON-STERILE, Rx only, Smith & nephew, Inc., Memphis TN 38116

This recall has been terminated (originally issued April 14, 2017).

Company
Smith & Nephew Inc
Recall Initiated
March 22, 2017
Posted
April 14, 2017
Terminated
June 3, 2022
Recall Number
Z-2036-2017
Quantity
12 units
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

This lot of left plates were laser etched and packaged as right plates.

Distribution

Internationally to the country of : India

Lot / Code Info

Lot 16DM15958

Root Cause

Mixed-up of materials/components

Action Taken

Smith & Nephew initiated their recall by email on March 22, 2017. Consignees are being requested to return all affected product in their possession. Distributors were instructed to notify their customers and ensure that the recall instructions were followed. For further questions, please call : 1(800) 821-5700.