Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated

This recall has been terminated (originally issued April 21, 2017).

Company
Stryker Communications
Recall Initiated
April 21, 2017
Terminated
September 20, 2017
Recall Number
Z-2248-2017
Quantity
57 units
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.

Distribution

Worldwide Distribution - US and Canada

Lot / Code Info

Serial Numbers: 7735170-X68377, 7735170-X68397, 7735170-X14846, 7735170-X15688, 7735170-X14492, 7735170-X14495, 7735170-X14497, 7735170-X14501, 7735170-X14502, 7735170-X14504, 7735170-X14506, 7735170-X14507, 7735170-X14508, 7735170-X14510, 7735170-X68383, 7735170-X68384, 7735170-X68385, 7735170-X68386, 7735170-X68393, 7735170-X68394, 7735170-X68395, 7735170-X68396, 7735170-X68392, 7735170-X68400, 7735170-X68401, 7735170-X15607, 7735170-X15608, 7735170-X15609, 7735170-X15610, 7735170-X15611, 7735170-X15612, 7735170-X15613, 7735170-X15614, 7735170-X15615

Root Cause

Component design/selection

Action Taken

Stryker Communications sent an Urgent Medical Device Recall letter to all affected customers. A Stryker Representative will contact their facility to schedule a service visit. to replace the adapter assembly to resolve the tolerance issue affected units. Customers with questions were instructed to Stryker Technical Support at 800-243-5135 or comm.techservice@stryker.com. For questions regarding this recall call 972-834-8656.