Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Image processing system.

This recall has been terminated (originally issued September 1, 2016).

Company
Merge Healthcare, Inc.
Recall Initiated
February 9, 2016
Posted
September 1, 2016
Terminated
August 24, 2017
Recall Number
Z-2709-2016
Quantity
116 customers potentially have the affected versions
Firm Location
Hartland, WI
Official Source
View on FDA website ↗

Reason for Recall

A report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR).

Distribution

Nationwide Distribution

Lot / Code Info

Versions 10.1 and 10.1.1

Root Cause

Software design

Action Taken

MERGE sent an Urgent Medical Device Recall letter dated February 9, 2016, to all affected customers via email or via certified mail if they did not have an e-mail address for the customer. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. In addition, the customer is requested to call Customer Service or send them an email to schedule the upgrade. If the customer chooses not to upgrade, they are requested to complete the attached form or go online and complete it. For further questions, please call (800) 724-5970 ext. 8027.