RecallDepth

Vertebral Motion Analyzer (VMA) Version 2.3.252. VMA software is a quantitative imaging software application intended to be used to process digital image files.

This recall is currently active, issued January 11, 2017. It was issued by Ortho Kinematics, Inc.

Company
Ortho Kinematics, Inc
Recall Initiated
January 11, 2017
Recall Number
Z-2882-2018
Quantity
1
Firm Location
West Lake Hills, TX
Official Source
View on FDA website ↗

Reason for Recall

Ortho Kinematics Inc. sent a Notice of Correction to Released Testing Results, Radiological Read Report for the Vertebral Motion Analyzer (VMA) test because it contained an error. The error occurred due to a software bug that has been corrected.

Distribution

US Nationwide Distribution

Lot / Code Info

UDI: 00868579000209

Root Cause

Software Manufacturing/Software Deployment

Action Taken

Ortho Kinematics Inc sent notification to customers on 1/11/2017 via Letter and electronic notification through web app. The company deployed a software upgrade (VMA version 2.3.253) on January 11, 2017.

More recalls by Ortho Kinematics, Inc

View all recalls by this company →