RecallDepth

ACUSON S Family ultrasound systems utilizing the 18L6 HD and/or 6C1 HD biopsy guidelines 6C1HD - model numbers: 10442008 6C1HD Needle Guide, S3000; 10440586 6C1 HD Biopsy Guide, S2000 18L6 - model numbers: 10442007 18L6 Biopsy Guide, S3000 10044200 S2000 Biopsy Starter Kit, 18L6 HD 10855189 18L6 HD Biopsy Starter Kit, S1000 Radiology: The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

This recall has been terminated (originally issued August 5, 2015).

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
June 29, 2015
Posted
August 5, 2015
Terminated
January 18, 2018
Recall Number
Z-2344-2015
Quantity
4140 total
Firm Location
Mountain View, CA
Official Source
View on FDA website ↗

Reason for Recall

When using the CIVCO biopsy attachments in combination with the ACUSON S Family on-screen guidelines, the biopsy needle may traverse outside the on-screen guidelines.

Distribution

Worldwide Distribution - US Nationwide and the countries of Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Brazil, Brunei, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Dem. Rep. Congo, Denmark, Dominican Rep., Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, P.R. China, Peru, Philippines, Poland, Portugal, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom, Venezuela and Vietnam.

Lot / Code Info

6C1HD - model numbers: 10442008 6C1HD Needle Guide, S3000; 10440586 6C1 HD Biopsy Guide, S2000; 18L6 model numbers: 10442007 18L6 Biopsy Guide, S3000 10044200 S2000 Biopsy Starter Kit, 18L6 HD 10855189 18L6 HD Biopsy Starter Kit, S1000  See Consignee list for all serial numbers.

Root Cause

Under Investigation by firm

Action Taken

Siemens sent a Customer Safety Advisory Notification letters to affected customers via certified mail with return receipt. The letter identified the affected product, problem, potential risk to patients and actions to be taken. Seimens instructed that this notice be shared with all personnel within your organization who need to be aware of this issue. For questions contact your Siemens service representative.

More recalls by Siemens Medical Solutions Usa, Inc.

View all recalls by this company →