Autoject E1 Fixed needle device, Model Number AJ1310, UPC Code 384701310010 The Autoject E1 is an injector used by patients to inject drugs as prescribed by their doctor.
- Company
- Owen Mumford Usa, Inc.
- Recall Initiated
- November 24, 2014
- Posted
- April 2, 2015
- Terminated
- July 31, 2017
- Recall Number
- Z-1366-2015
- Quantity
- 3,100
- Firm Location
- Marietta, GA
Reason for Recall
The syringe carrier is missing components: a damper and spring.
Distribution
US nationwide distribution in the states of AL, AR, CA, FL, GA, IL, MO, NC, OK, TN, and TX.
Lot / Code Info
Lot No. 5LW
Root Cause
Process control
Action Taken
Recall letters were issued to customers via FedEx. The letter instructed the customer to check their shelves for the product and specific lot number and to visually inspect (pictures included in letter of a correctly assembled device and one with the missing spring & damper) the device if they have the affected lot. A form was included which was to be filed out and returned.