Autoject EI, REF AJ1310
- Company
- Owen Mumford Usa, Inc.
- Recall Initiated
- July 17, 2019
- Recall Number
- Z-0263-2020
- Quantity
- 1510 units
- Firm Location
- Marietta, GA
Reason for Recall
There is a possible assembly error on Autoject EI, lot number V14. This lot was packed into two products: AJ 1310, lot number V14 and AN 3310VET (branded as Petfine Auto Injector), batches V14.V976 & V14.W107
Distribution
FL, LA, SC, CA, NJ, TX, NC, CO, NH, KY, NY, MO, PA
Lot / Code Info
Lot/Unit Number: V14
Root Cause
Employee error
Action Taken
The firm initiated the recall by letter on 07/17/2019. The consignee was directed to quarantine all affected product and contact the company for return. Distributors were also directed to cease distribution and contact their customers.