RecallDepth

MEVION S250, used for proton radiation therapy.

Company
Mevion Medical Systems, Inc.
Recall Initiated
November 10, 2014
Posted
December 12, 2014
Terminated
March 18, 2015
Recall Number
Z-0588-2015
Quantity
1
Firm Location
Littleton, MA

Reason for Recall

Software defect that causes an incorrect dose compensation function to be applied to the internal dose ionization chamber when either pressure or temperature sensor malfunction. This could result in an incorrect dose delivery of no more than 5%.

Distribution

US Distribution in the state of MO

Lot / Code Info

Serial Number: S250-0001

Root Cause

Software design

Action Taken

Mevion Medical Systems sent an Important Safety Notice letter to the affected customer. The letter identified the affected product, problem and actions to be taken. A software update addressing the issue will be provided to the customer on or about December 5, 2014. For questions or concerns, email lbouchet@mevion.com or call 978-540-1555.

More recalls by Mevion Medical Systems, Inc.

View all recalls by this company →