Hudson RCI Pediatric Anesthesia Breathing Circuits, Single Patient Use, Rx Only, Product Usage: To administer medical gases to a patient during anesthesia. It provides both an inhalation and exhalation route and may include a connector, adaptor, and Ypiece.
This recall has been terminated (originally issued September 30, 2014).
- Company
- Teleflex Medical
- Recall Initiated
- September 8, 2014
- Posted
- September 30, 2014
- Terminated
- August 16, 2016
- Recall Number
- Z-2714-2014
- Quantity
- 27,176 units
- Firm Location
- Durham, NC
- Official Source
- View on FDA website ↗
Reason for Recall
The pediatric breathing circuits can crack prior to and during use.
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, CA, FL, IL, KY, OH, NJ, NY, TX, LA, OR, SD, WA, including Puerto Rico, and the countries of Belgium, Canada, Chile, China, Colombia, Korea, Mexico, Peru, Venezuela and Zaire.
Lot / Code Info
Product Code: 313905, Lot #02F1300085; Product Code: 353900, Lot #02D1300584; Product Code: 353901, Lot #0241300618; Product Code: 353911, Lot #02G1300296, 02G1300861, 02F1300729; Product Code: 313901, Lot #02G1300088; Product Code: 353811, Lot #02E1300090; 02E1301383 & 02E1301867; Product Code: 353911, Lot #02F1300102 & 02F1300177; Product Code: 353900, Lot #02F1302029; Product Code: 353801, Lot #02E1300623, 02FF1300098, 02H1300172, 02H1300454, 02H1301128, 02F300180, 02E1301891, 02J1300876, 02H1301838, 02J1302412 & 02B1400324; Product Code: 353901, Lot #02E1301378; Product Code: 353900, Lot #02E1301892, 02F1300100 and Product Code: 353901, Lot #02F1302047, 02F1300101 & 2G1300879.
Root Cause
Device Design
Action Taken
Teleflex Medical sent an Urgent Medical Device Recall dated September 8, 2014, to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to immediately discontinue use, quarantine any products with the product codes and lot numbers provided in the notification, return any affected product, complete and return the enclosed Recall Acknowledgement Form and fax it to the attention of Customer Service. For questions contact your local sales representative or Customer Service at 1-866-246-6990.