Aquapak Adaptor, 028 NEB, INTL, Product No. 403128; Aquapak Adaptor, 033 NEB, INTL, Product No. 403133; Aqaupak Nebulizer Adaptor 028, Sterile, Shelfpak, Product No. 031-28 & 031-28F, Aqaupak Nebulizer Adaptor 033, Sterile, Shelfpak, Product No. 031-33; Aqaupak Nebulizer Adaptor 033, Sterile, Japanese, Product No. 031-33J. Used in anesthesiology.
This recall has been terminated (originally issued December 12, 2014).
- Company
- Teleflex Medical
- Recall Initiated
- September 3, 2014
- Posted
- December 12, 2014
- Terminated
- August 14, 2017
- Recall Number
- Z-0504-2015
- Quantity
- 3,193,548 ea
- Firm Location
- Durham, NC
- Official Source
- View on FDA website ↗
Reason for Recall
Possibility that the nebulizer adaptor packaging may have open or weak seals which may affect the sterility of the adaptor.
Distribution
Worldwide Distribution-US (nationwide) including Puerto Rico and the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, WY, and the countries of United Arab Emirates, Australia, Belgium, Brazil, Canada, Chile, China, France, Great Britain, Japan, India, Indonesia, Israel, South Korea, Mexico, Philippines, Saudi Arabia, Singapore, Taiwan, Thailand, Uruguay, and Zambia.
Lot / Code Info
Product No. 043128, Lot # 02E1200719, 02E1200720, 02E1201281, 02F1200964, 02K1201946, 02E1201808, 02E1203097, 02G1300256, 02G1301048, & 02M1201691; Product No. 403133, Lot # 02E1201282, 02K1201948, 02F1200965, 02L1201194, 02M1200230, & 02F1302063; Product No. 031-28, Lot # 02D1202234, 02A1202383, 02A1202382, 02B1201069, 02D1200247, 02D1200248, 02D1201944, 02D1202274, 02D1202733, 02D1202236, 02D1202734, 02D1202735, 02D1202233, 02G1200664, 02G1200678, 02G1201016, 02G1201698, 02F1200961, 02G1200676, 02M1200585, 02L1200927, 02M1200749, 02M1201693, 02A1300194, 02A1300457, 02D1301655, 02D1301191, 02D1300864, 02D1301192, 02D1302730, 02F1301742, 02E1302765, 02F1300739, 02F1302062, & 02D1200249; Product No. 031-28F, Lot # 02E1200717, 02K1201947, 02F1200962, & 02E1202197; Product No. 031-33, Lot # 02E1200718, 02A1202384, 02E1202198, 02F1200963, 02G1200666, 02J1202240, 02M1200584, 02M1200748, 02M1201771, 02D1302731, 02A1300458, 02B1300718, & 02M1201692; Product No. 031-33J, Lot # 02D1202258, 02D1202260, 02D1202259, 02D1202261, 02E1201563, 02E1201279, 02E1203866, 02M1200747, 02B1300717, 02B1301203, 02E1301554, 02B120342, 02J1201473, 02J1202736, 02F1201179, & 02M1200057.
Root Cause
Packaging change control
Action Taken
Customers were notified of the recall by letter on September 3, 2014. The letter requested that they discontinue distribution and quarantine the products and return the products to Teleflex. The letter also requested a sub-recall.