Edwards Lifesciences Fem-Flex II Arterial and femoral Cannulae with Duraflo Coating, arterial cannula sizes 16, 18 and 20 French, Models DFEMII016AS, DFEMII018AS, DFEMII020AS, DIIFEMII016A, DIIFEMII018A, and DIIFEMII020A.
This recall has been terminated (originally issued April 16, 2014).
- Company
- Edwards Lifesciences, Llc
- Recall Initiated
- April 9, 2014
- Posted
- April 16, 2014
- Terminated
- December 28, 2016
- Recall Number
- Z-1467-2014
- Quantity
- 28,885
- Firm Location
- Draper, UT
- Official Source
- View on FDA website ↗
Reason for Recall
Potential for separation of dilator tips.
Distribution
Nationwide, including foreign, govt/VA/Military consignees.
Lot / Code Info
DFEMII016AS - 59036042, 59213298, 59258406, 59578907; DFEMII018AS - 59176126, 59249371, 59271003, 59369527, 59374074, 59432937, 59468460, 59484133, 59585854, 59589948, 59645287 DFEMII020AS - 59036043, 59113935, 59213299, 59258407, 59320308, 59369528, 59432938, 59468461, 59484119 DIIFEMII016A- 59025196, 59069368, 59115263, 59134580, 59154519, 59218668, 59240886, 59248906, 59304339, 59342282, 59383853, 59409789, 59453487, 59473491, 59491809, 59523710, 59528324, 59555055, 59589915, 59625359, 59645284, 59651762, 59680133, 59691448 DIIFEMII018A - 59030249, 59060764, 59072416, 59093901, 59116037, 59129577, 59149668, 59176262, 59218745, 59240935, 59248907, 59304334, 59337174, 59350865, 59394866, 59431915, 59468375, 59493314, 59528325, 59555040, 59577156, 59599140, 59609142, 59625362, 59651760, 59651808, 59680241 DIIFEMII020A - 59043574, 59069369, 59082560, 59103193, 59124011, 59134578, 59149667, 59176263, 59191993, 59218735, 59229561, 59240934, 59266786, 59299945, 59318926, 59319458, 59337178, 59350871, 59404487, 59437572, 59473387, 59484151, 59516163, 59531221, 59558959, 59563837, 59585992, 59609143, 59620175, 59642447, 59651767, 59651823, 59662388, 59680247.
Root Cause
Nonconforming Material/Component
Action Taken
Customers were notified via letter on 4/4/14 to return affected products.