3M" Steri-Vac" Gas Sterilizer/Aerators, Model #'s 4XL,5XL, 8XL The 3M" Steri-Vac" Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or moisture-sensitive devices. This gas sterilizer/aerator is intended for INDOOR USE ONLY.

Company
3M Company Health Care Business 3M Center
Recall Initiated
March 17, 2014
Posted
April 2, 2014
Terminated
January 22, 2015
Recall Number
Z-1318-2014
Quantity
73 US, 346 OUS
Firm Location
Saint Paul, MN

Reason for Recall

Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installation involving 3M" Steri-Vac" Gas Sterilizers. The cause was traced to a rare situation involving a failure of the EO cartridge puncture assembly. This assembly was only in production during the period November 10, 2006 through December 14, 2007 and in service during the period February 7, 20

Distribution

Worldwide Distribution - USA (nationwide) Distribution including the states of AL, AZ, CA,CO, FL, GA, HI, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, NE, NJ, NY, NC, ND, OH, OK,OA, SC, TX, WA, WV and WI. and the countries of COLOMBIA, VENEZUELA, UAE, THAILAND, GERMANY, CHINA, KOREA, CANADA, JAPAN, RUSSIA, INDIA, POLAND, PERU, AUSTRIA, ISRAEL, INDONESIA, SPAIN, CHILE, MALAYSIA, SWITZERLAND, EL SALVADOR, ECUADOR, LEBANON, MEXICO, ARGENTINA, BULGARIA, TAIWAN, TURKEY,ROMANIA, GUATEMALA and NEW ZEALAND.

Lot / Code Info

Model SKU Starting S/N Ending S/N 5XL 70-2007-2303-2 721115 721309 5XL 70-2007-2304-0 820174 820201 5XL 70-2007-2575-5 780066 780102 8XL 70-2006-8493-7 350136 350268 8XL 70-2006-8498-5 450016 450027  4XL 523122, 522205, 522502, 522287, 522655.

Root Cause

Device Design

Action Taken

3M sent an "Urgent Medical Device Field Correction" letter dated March 6, 2014. The letter was sent to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. You are being notified because our records indicate you received a unit that contains the cartridge puncture assembly that is the subject of this field correction. 3M will contact you to schedule an appointment for a field correction on any unit in the range of serial numbers provided in the table. 3M's goal is to service the affected units with minimal disruption to your facility. Units manufactured outside of the dates November 10, 2006 to December 14, 2007 are not affected. .For Further Information contact 3M Health Care Service Line at 1-800-228-3957 or your local 3M representative.