Alere Cholestech LDX ALT AST Test Cassette, Model #12-788. For the in vitro quantitative determination of alanine aminotransferase (ALT) and aspartate aminotransferase (AST).
- Company
- Alere San Diego, Inc.
- Recall Initiated
- March 6, 2013
- Posted
- April 17, 2013
- Terminated
- August 13, 2013
- Recall Number
- Z-1139-2013
- Quantity
- 38,571 kits
- Firm Location
- San Diego, CA
Reason for Recall
The recall was initiated because Alere San Diego has an update for the Alere Cholestech LDX ALT AST cassette, and it has the potential for humidity changes to impact results for certain analytes on the Alere Cholestech LDX System.
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, South Africa, Netherlands, Italy, Brazil, Singapore, Russia, Hong Kong, Argentina, South Korea, Philippines, El Salvador, Indonesia, and Trinidad.
Lot / Code Info
Lot/Unit Codes: 274351A, 274351B, 274352B, 274353A, 290243A, 290244A, 290641A, 290641B, 290642A, 290643A, 290643B, 290644A, 290644B, 290645A, 290645B, 290646A, 290647A, 290648A, 293669A, 293670A, 293671A, 293672A, 293673A, 296635A, 296636A, 296637A, 296638A, 296639A, 296639B, 296640A, 296641A, 296642A, 297311B, 297313A, 297313B.
Root Cause
Other
Action Taken
Alere sent an "URGENT MEDICAL DEVICE RECALL UPDATE" letter dated March 6, 2013 to all affected customers. The letter informed the customers of the problem identified and the actions to be taken. Customers with any questions about the recall notification, are instructed to contact Alere San Diego 9975 Summers Ridge Rd, San Diego, CA 92121; (877) 308-8289; Fax: (858) 805-8457. Customers are instructed to complete the customer response form even if they do not have any involved product.