RecallDepth

Triage Drugs of Abuse Plus TCA 25 Test Kit, Model Numbers 92000, 92590, 92591; 25 individually pouched devices, 1 bottle of Wash Solution, 1 Pipette and 50 Pipette Tips in a labeled kit box. Product Usage: The Alere Triage Drugs of Abuse Panel plus Tricyclic Antidepressants uses eight discrete competitive immunoassay procedures for the simultaneous detection and identification of the major drugs and/or the urinary metabolites of eight different drug classes. The use of monoclonal antibodies that are specific for the drugs and/or metabolites of the eight drug classes ensures a high degree of sensitivity and specificity.

Company
Alere San Diego, Inc.
Recall Initiated
November 14, 2017
Recall Number
Z-0418-2018
Quantity
2940 units
Firm Location
San Diego, CA

Reason for Recall

The recalled lots have demonstrated unexpected false positive THC results.

Distribution

Worldwide distribution - United States Nationwide and the countries of Japan, Canada

Lot / Code Info

Lots 403111, 406741, 396588, 398577, 396587

Root Cause

Nonconforming Material/Component

Action Taken

All primary consignees will be contacted in letter provided via fax, email or direct mail. End users will be instructed to discard any remaining inventory they may have of the affected lots. Alere will provide credit to end users for any remaining unused inventory.

More recalls by Alere San Diego, Inc.