Indus Invue Screw Caddy, Model: SI70097 The caddy holds, in an organized fashion within the case, a variety of shapes and sizes of plates and screws used during fixation of anterior cervical plates to provide stabilization during the development of spine fusions. There is an array of 108 holes that store 3 types of screws. O4.2mm Self-Drilling, O4.2mm Self-Tapping, and O4.5mm Self-Tapping screws which are separated by type and screw length. The caddy has features to hold the screws in place. Recall expanded on 8/21/2013 to include additional lots of this product.

This recall has been terminated (originally issued August 8, 2013).

Company
Spinefrontier, Inc.
Recall Initiated
June 21, 2013
Posted
August 8, 2013
Terminated
January 11, 2016
Recall Number
Z-1933-2013
Quantity
13
Firm Location
Beverly, MA
Official Source
View on FDA website ↗

Reason for Recall

Difficulty removing screws from the Invue caddies.

Distribution

Worldwide Distribution - USA, including the states of CT, VA, MA, FL, TX, CO and MS. and the country of Jamaica.

Lot / Code Info

Lot Numbers: 7140.014; 6486.008  Recall expanded on 8/21/2013 to include lots: 10754.011; 6417.005R, 7140.005, 6486.007; 6486.007R

Root Cause

Device Design

Action Taken

SpineFrontier sent a Urgent Invue Caddy Advisory letter dated July 10, 2013, to all affected customers. Customers were advised the removal of the Invue Screw Caddy, SI70097 from the field for modification. Customers were asked to return the parts to SpineFrontier and to complete an Acknowledgement Form indicating that they have read the notice and will initiate the immediate return of the parts noted above. Please feel free to contact us at (978) 232-3990. On 8/21/2013 the recall was expanded.