Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A; Invue Lid: SI70095A; S-Lift Lid: SI50070A; S-Lift Base: SI50067A; Arena-C Lid: 11-72013D, 11-72014D; Generics Lid: 12-99005A. System cases are designed to hold and secure all instrumentation to be used with a surgical system.

This recall has been terminated (originally issued August 13, 2013).

Company
Spinefrontier, Inc.
Recall Initiated
April 22, 2013
Posted
August 13, 2013
Terminated
November 6, 2015
Recall Number
Z-1958-2013
Quantity
258 units (30 in distribution at the time of the recall)
Firm Location
Beverly, MA
Official Source
View on FDA website ↗

Reason for Recall

Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles.

Distribution

Worldwide Distribution-USA including the states of FL, OH, MA, and TX, and the countries of Mexico, and Dominican Republic.

Lot / Code Info

Inspan Lid: Part Number: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A, Lot Numbers: 7197.024, 011649, 011498; Invue Lid: SI70095A, Lot Number: 6250.009-R; S-Lift Lid: SI50070A, Lot Number: 7201.008; S-Lift Base: SI50067A; Lot Numbers: 7183.007R, 7201.007; Arena-C Lid: 11-72013D, 11-72014D, Lot Numbers: 011969-001, 011969-002; Generics Lid: 12-99005A, Lot Number: 8372.001.

Root Cause

Labeling design

Action Taken

On 2/15/2013, SpineFrontier emailed distributors informing them that they needed specific parts returned to work towards completing the conversion of product graphics.