Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.

This recall has been terminated (originally issued April 8, 2013).

Company
Toshiba American Medical Systems Inc
Recall Initiated
November 14, 2012
Posted
April 8, 2013
Terminated
April 8, 2013
Recall Number
Z-1068-2013
Quantity
1 unit
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.

Distribution

Distributed in the state of MD.

Lot / Code Info

Serial No: 1CB1262006

Root Cause

Process design

Action Taken

Toshiba American Medical Systems Inc. sent a Urgent Medical Device Correction letter dated November 14, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customer was informed a New software to prevent the occurrence of this issue will be installed on your system. Toshiba Service Representative will contact you for an appointment to install the New Software on your Aquilion One System. Please fax the return reply form to (877) 349-3054. We thank you for your urgent attention to this matter. If you have any further questions regarding this letter please call (800) 521-1968.