i-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540

This recall has been terminated (originally issued July 9, 2012).

Company
Abbott Point Of Care Inc.
Recall Initiated
September 8, 2011
Posted
July 9, 2012
Terminated
January 28, 2013
Recall Number
Z-1976-2012
Quantity
6000-0000-650 units
Firm Location
Princeton, NJ
Official Source
View on FDA website ↗

Reason for Recall

Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.

Distribution

Worldwide Distribution - USA (nationwide)

Lot / Code Info

i-Stat1 Abaxis Analyzer  6000-0000  All lots

Root Cause

Device Design

Action Taken

Abbott Point Care sent an Urgent Product Recall letter dated September 2011 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to confirm the number of affected product at their facility that were using the Old Battery Carrier. Fill out the affected Buesiness Reply Card and submit their facility information as per the instructions on the card. Upon the receipt of the New Battery Carriers, replace the Old Battery Center with the New Battery Center. Customers were instructed to immediately dispose of all Old Battery Carriers per local regulations. If customers forward the affected product to another facility, they should provide a copy of the recall letter to them and advise them to replace the carriers if required. For any questions call 800-366-8020, Option 1 or your Abbott Point of Care representative. For questions regarding this recall call 609-454-9272.

More recalls by Abbott Point Of Care Inc.

View all recalls by this company →