Touch Panel/Control Unit Carrier of the Swingarm M820/M844 C19, article no. 10448194MO; a component of the Leica M820 C19 Surgical Microscope, article 10448196, and the Leica M844 C19 Surgical Microscope, article 10448197; ceiling mounted surgical microscopes; Leica Microsystems (Schweiz) AG, Max Schmidheiny-Strasse 201, CH-9425, Heerbrugg, Switzerland;; The intended use of the microscope is for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

This recall has been terminated (originally issued March 21, 2012).

Company
Leica Microsystems, Inc.
Recall Initiated
February 27, 2012
Posted
March 21, 2012
Terminated
May 18, 2012
Recall Number
Z-1269-2012
Quantity
33 units
Firm Location
Buffalo Grove, IL
Official Source
View on FDA website ↗

Reason for Recall

The screws, which lock the bottom plate to the spindle, pull through allowing the carrier arm with touch panel/control unit of the surgical microscope to disconnect, which potentially could result in them falling.

Distribution

Worldwide distribution: USA (nationwide) including states of: California, Connecticut, Kansas, Missouri, New York, North Dakota, Ohio, Pennsylvania, Texas and Utah; and countries of: Japan and South Africa,

Lot / Code Info

article 10448196: serial numbers 17091004, 20911002, 26110801, 31207001, 31207002, 61006001, 61006002, 91208002, 101208001, 101210002, 140508002, 151209002, 160806002, 260811001, 261108001 and 261108002;  article 10448197: serial numbers 11010001, 50510002, 81007003, 110507002, 130707001, 130707003, 160910008, 180607001, 230909001, 270407006 and 300307001

Root Cause

Device Design

Action Taken

The firm, Leica Microsystems, sent a "Medical Device Correction" letter dated February 27, 2012 to the affected customers via e-mail on the same date. The letter described the product, problem and action to be taken. The customers were informed that the Leica M820 and M844 C19 surgical microscopes require upgrading. An upgrade kit will be provide for replacement of the flange and axis of the touch panel/control unit carrier. The Customers were instructed to NOT USE the affected Leica surgical microscope until the unit has been upgraded by their Leica Microsystems representative and to complete the attached Acknowledgement Form, acknowledging the receipt and understanding of the letter, listing the serial numbers of the unit at their facility and fax it back to Leica at 1-847-236-3747. If they have any questions, contact Director, Global Product Quality Management at 1-847-317-7209.