Animas ezManager Max Diabetes Management System version 2.0.9

This recall has been terminated (originally issued December 3, 2009).

Company
Animas Corporation
Recall Initiated
November 2, 2009
Posted
December 3, 2009
Terminated
October 27, 2010
Recall Number
Z-0481-2010
Quantity
20,648 units
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

Printed hard copy of results is not the same as the system screens- carbohydrate to insulin ratio incorrect.

Distribution

Nationwide distribution, and Canada, Cayman Islands, and Denmark.

Lot / Code Info

version 2.0.9

Root Cause

Software design

Action Taken

Animas issued an Urgent Medical Device Correction letter dated 11/2/09 to the pump user and another letter to the health care professional that prescribed the pump. The letters explain the issue. The patient letter is accompanied with a CD that contains the upgraded software and instructions to follow.