Oxygen Sensor #5804, Distributed by: Teleflex Medical, Research Triangle Park, NC 27709

This recall has been terminated (originally issued December 29, 2009).

Company
Teleflex Medical
Recall Initiated
November 25, 2009
Posted
December 29, 2009
Terminated
April 7, 2011
Recall Number
Z-0585-2010
Quantity
51 eaches
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

Sensor causes the oxygen monitor to display prematurely a Low Sensor warning which indicates an early maintenance requirement for the instrument. This may result in interrupted treatment.

Distribution

FL, KY, MI, OH, PA, SD, TN, Australia and Canada.

Lot / Code Info

Catalog number: 5804, Lot numbers 906018 and 906003.

Root Cause

Vendor change control

Action Taken

Consignees were notified via next day delivery on or about 11/25/2009. Each customer received a letter requesting acknowledgement of the recall and a request to return the affected product. Each consignee that is a distributor was instructed to forward the letter to their consignees to retrieve relevant product from those locations. A second and third mailing to non-responding customers on the consignee list will be made, or as an alternative, the third contact may be by phone, e-mail, or fax. Affected products received by Teleflex Medical will be destroyed or reworked and replacement product or credit will be provided to the consignee. Records of contacts and of destruction or rework will be maintained. Questions are directed to the company at 1-866-246-6990.