GE Healthcare, CT Perfusion 4 on Advantage Workstation, CT and PET/CT consoles. Image analysis software package that allows the user to produce image data and to generate information with regard to changes in image intensity over time supporting CT Perfusion images.

This recall has been terminated (originally issued August 13, 2009).

Company
Ge Medical Systems, Llc
Recall Initiated
May 4, 2009
Posted
August 13, 2009
Terminated
May 16, 2012
Recall Number
Z-1720-2009
Quantity
444
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

GE Healthcare has recently become aware of the need to adjust default settings associated with the use of CT Perfusion 4 on the Advantage Workstation, ST and PET/CT consoles that may impact patient safety. The default parameter settings in CT Perfusion 4 may not be optimal in the case of data acquired with longer time sampling intervals such as axial scans with inter scan delay (ISD) greater th

Distribution

Worldwide Distribution -- Including states of AZ, CA, CT, DC, FL, HI, IL, IA, MA, MI, NJ, NY, NC, OK, PA, PR, TX, WA, WV and WI and countries of ARGENTINA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONDURAS, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PAKISTAN, POLAND, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED KINGDOM and VENEZUELA.

Lot / Code Info

Serial Numbers: 00000152741GE5, 00000152737GE3, 00000134590GE9, 00000119522GE1, 00000119523GE9, 00000146655GE6, 00000146668GE9, 00000146669GE7, 00000119519GE7, 00000145765GE4, 00000152738GE1, 00000121182GE0, 00000122793GE3, 00000119517GE1, 00000139086GE3, 00000141801GE1, 00000141802GE9, 00000141803GE7, 00000141804GE5, 00000141805GE2, 00000119521GE3, 00000118133GE8, 00000118134GE6, 00000118135GE3, 00000121180GE4, 00000141807GE8, 00000119516GE3, 00000123926GE8, 00000123931GE8, 00000122794GE1, 00000127524GE7, 00000127525GE4, 00000129247GE3, 00000134588GE3, 00000123927GE6, 00000123929GE2, 00000119520GE5, 00000122797GE4, 00000122795GE8, 00000123930GE0, 00000127522GE1, 00000127523GE9, 00000129243GE2, 00000127520GE5, 00000127521GE3, 00000127526GE2, 00000139088GE9, 00000129246GE5, 00000145645GE8, 00000141808GE6, 00000141811GE0, 00000151783GE8, 100995344, 00000145642GE5, 00000145647GE4, 00000118132GE0, 00000122799GE0, 00000122800GE6, 00000123928GE4, 00000127519GE7, 00000152735GE7, 00000129244GE0, 00000129245GE7, 00000129248GE1, 00000129249GE9, 00000129250GE7, 00000129251GE5, 00000134584GE2, 00000134585GE9, 00000134586GE7, 00000134587GE5, 00000145649GE0, 00000152742GE3, 00000122798GE2, 100995241, 100995217, 100995259, 80521349, 100995302, 100995275, 100995276, 100995431, 100995326, 100995412, 100995322, 80515302, 80515303, 80523248, 100995352, 100995298, 100995359, 100995306, 100995268, 100995354, 0CC08041501002, 0CC08041501003, 100995225, 80528332, 80521354, 100995279, 100995280, 100995281, 100995282, 100995283, 100995297, 100995305, 100995308, 100995288, 100995340, 100995355, 100995361, 100995385, 100995356, 100995450, 100995227, 80528736, 0CC08041501009, 80528724, 100995242, 100995351, 100995211, 100995212, 100995383, 80527216, 80527225, 80515730, 80528335, 80515721, 80521344, 80528730, 80528741, 80521352, 80521350, 80523251, 80521348, 100995222, 80515722, 100995339, 00000134592GE5, 00000134591GE7, 80505225, 80505229, 0CC08041501005, 80528328, 80528336, 80529188, 80521353, 100995224, 100995218, 100995221, 100995360, 100995240, 100995266, 100995267, 100995274, 80528329, 80528327, 00000121178GE8, 00000141809GE4, 00000152739GE9, 00000155218GE1, 00000121179GE6, 0CC08041501008, 100995239, 100995294, 100995327, 80528734, 100995366, 80515727, 80528322, 80521355, 100995278, 100995328, 100995331, 100995396, 100995411, 100995420, 00000146658GE0, 80528723, 80528324, 80505223, 80505226, 80505228, 100995220, 80528325, 0CC08041501007, 80527223, 80515296, 80528731, 80527219, 100995408, 0CC08041501010, 80528733, 80528735, 80528732, 80529189, 80521356, 100995311, 100995364, 100995219, 100995271, 100995292, 100995338, 100995334, 100995336, 100995286, 100995350, 100995379, 100995317, 100995318, 100995413, 100995449, 100995375, 80523250, 80515297, 80521351, 80528326, 0CC08041501004, 100995205, 80521343, 0CC08041501001, 100995270, 80505230, 80527218, 100995265, 100995300, 100995400, 100995426, 100995433, 80527224, 100995330, 100995341, 100995378, 100995291, 80528323, 0CC08041501006, 80505227, 100995377, 100995295, 100995272, 100995312, 100995368, 100995284, 100995349, 100995245, 80528330, 100995289, 100995255, 100995256, 100995374, 100995406, 100995319, 100995415, 100995409, 100995401, 100995332, 100995325, 100995419, 100995405, 100995407, 100995427, 100995429, 100995417, 80523253, 80528333, 00000145766GE2, 00000145644GE1, 80515298, 100995260, 0CC08040101001, 100995249, 100995388, 100995425, 100995229, 80523247, 100995209, 80528337, 100995296, 80515724, 80528339, 80505231, 80528739, 80515723, 80528340, 80515725, 80529195, 100995244, 80523249, 80515728, 80515729, 100995348, 100995370, 100995410, 100995376, 80529190, 80529191, 80529192, 80515726, 80529186, 80523246, 80528338, 80527222, 80529194, 80529193, 80528725, 80521346, 80528331, 80528334, 80528321, 80521347, 100995250, 100995251, 100995252, 100995254, 100995342, 000005189396-2, 100995223, 80523252, 100995215, 100995236, 100995231, 100995233, 100995232, 100995234, 100995237, 100995230, 100995235, 100995247, 100995304, 100995269, 100995273, 100995258, 100995285, 100995337, 100995287, 100995246, 100995314, 100995264, 100995387, 100995343, 100995365, 100995390, 100995392, 100995394, 100995391, 100995393, 100995384, 100995329, 100995404, 100995395, 100995402, 80528740, 80528737, 80527217, 80527221, 80528726, 80528727, 80528728, 100995316, 100995345, 100995353, 100995367, 80505224, 80505232, 100995207, 00000146661GE4, 00000144815GE8, 00000146660GE6, 00000146662GE2, 100995418, 100995441, 100995416, 00000146664GE8, 00000146666GE3, 00000146665GE5, 00000146659GE8, 100995238, 80529187, 100995208, 80523254, 100995214, 100995213, 100995206, 80515301, 80515305, 80527220, 100995253, 80523255, 100995210, 80515299, 80515304, 100995226, 100995257, 100995323, 100995228, 100995248, 100995262, 100995362, 100995301, 100995307, 100995309, 100995310, 100995299, 100995346, 100995403, 100995373, 100995363, 100995389, 100995422, 100995380, 100995381, 100995434, 00000145763GE9, 00000134589GE1, 00000121181GE2, 00000119518GE9, 00000151784GE6, 00000134583GE4, 00000141810GE2, 00000145641GE7, 00000145640GE9, 00000145768GE8, 00000145764GE7, 00000145767GE0, 00000151782GE0, 00000139087GE1, 00000145643GE3, 00000151780GE4, 00000151781GE2, 00000152740GE7, 00000152736GE5, 00000145646GE6, 00000145648GE2, 100995263, 100995243, 100995261, 100995421, 100995436, 100995435, 100995423, 100995424, 00000118131GE2, 00000141806GE0 and 00000118136GE1.

Root Cause

Software design

Action Taken

GE Healthcare issued an "Urgent Medical Device Correction" letter dated April 23, 2009. The letter was addressed to Hospital Administrators / Risk managers and Radiology department mangers. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information instructing the user to follow the enclosed User Guide Addendum to mitigate further problems. For further questions, contact GE Healthcare at 1-800-437-1171 (US), 0120-055-919 (Japan); for other countries, contact your local GE Healthcare field service representative.