Bausch & Lomb Millennium Phacoemulsification Needles, Needle Tapered Hub 15 Degree, REF DP8115, 6 needles per pouch, nonsterile.

This recall has been terminated (originally issued September 16, 2008).

Company
Bausch & Lomb Inc
Recall Initiated
March 26, 2008
Posted
September 16, 2008
Terminated
January 21, 2010
Recall Number
Z-2044-2008
Quantity
Total of all sizes - 4,055/6-needle packages
Firm Location
Rochester, NY
Official Source
View on FDA website ↗

Reason for Recall

The directions for use are printed with the incorrect symbols to indicate the product is packaged as sterile and nonreusable

Distribution

Distribution was made nationwide direct to medical facilities, which included some government accounts. Foreign distribution was made to Australia, Brazil, Bulgaria, Canada, China, Colombia, Cyprus, England, France, Germany, India, Iran, Ireland, Israel, Italy, Jordan, Korea, Lebanon.,The Netherlands, Poland, Portugal, Russia, Saudi Arabia, South Africa, Spain, Thailand, Turkey, and United Arab Emerites.

Lot / Code Info

Lot number 3168901

Root Cause

Labeling Change Control

Action Taken

The recalling firm issued letters dated 3/26/08 through a contracted recall notification company who issued the letters via UPS return receipt. A confirmation form was enclosed which was to be completed and returned. If you have questions, contact Customer Service at 1-800-338-2020.