Cemented Trapezoid Tibial Tray Sz. 4F/4T, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.

This recall has been terminated (originally issued December 23, 2004).

Company
Exactech, Inc.
Recall Initiated
November 11, 2004
Posted
December 23, 2004
Terminated
May 3, 2012
Recall Number
Z-0347-05
Quantity
21 implants
Firm Location
Gainesville, FL
Official Source
View on FDA website ↗

Reason for Recall

Implant was identified incorrectly at the time of manufacture and was then labeled and distributed with the incorrect size information.

Distribution

Distributed to New Jersey, Florida, Minnesota, New York, Maryland, Texas and the United Kingdom

Lot / Code Info

Serial numbers 0638851-0638871

Root Cause

Other

Action Taken

A recall letter was sent to all customers on November, 2, 2004 recalling all 21 units. These implants will be quarantined and then returned to the supplier for disposal when all 21 implants are received.