Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190.

This recall has been terminated (originally issued June 9, 2007).

Company
Roche Diagnostics Corp.
Recall Initiated
April 30, 2007
Posted
June 9, 2007
Terminated
December 17, 2007
Recall Number
Z-0920-2007
Quantity
992
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.

Distribution

Nationwide.

Lot / Code Info

Lots 144107, exp. 9/30/2007; 144166, exp. 1/31/2008, and 144208, exp. 6/30/2008

Root Cause

Other

Action Taken

Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.