OPMI VARIO 700 Surgical Microscope. Manufactured by: Carl Zeiss Meditec AG, Oberkochen, Germany. Provide a magnified view of the surgical field.

Company
Carl Zeiss Meditec Inc
Recall Initiated
September 26, 2011
Posted
October 24, 2011
Terminated
November 4, 2011
Recall Number
Z-0105-2012
Quantity
3
Firm Location
Dublin, CA

Reason for Recall

An incorrect nut and bolt set was used in the axis three (3) spring assembly of the OPMI VARIO 700 surgical microscope distributed to three US locations.

Distribution

USA (nationwide) including the states of FL, GA, and LA .

Lot / Code Info

Serial numbers: 6636101067; 6636101068; 6636101070.

Root Cause

Nonconforming Material/Component

Action Taken

Carl Zeiss Meditec had Field Service Engineers directly contact all affected customers. The Field Service Engineers will replace the defective component. Questions regarding this Field Corrective Action should be directed to the Director of Clinical/Regulatory Affairs at 925-557-4616.