Amico E-Cylinder Canister Holder, distributed by Stryker Communications, Flower Mound, TX 75028. Canister holder is used to hang E-Cylinders (pressurized gas canisters) from a boom. Product typically used in the surgical setting.

This recall has been terminated (originally issued July 13, 2009).

Company
Stryker Communications Corp
Recall Initiated
May 19, 2009
Posted
July 13, 2009
Terminated
April 26, 2010
Recall Number
Z-1569-2009
Quantity
66 units.
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

Design flaw on the E-Cylinder canister holder may result in early fatigue and failure of the bracket screws, which could cause the cage to become detached from its mount and fall.

Distribution

Nationwide and Canada.

Lot / Code Info

Catalog/Product Number: 0682-001-111 and 0682-000-449.  All E-Cylinder canister holders shipped between 01 JAN 05 and 27 SEP 06.

Root Cause

Device Design

Action Taken

An "Urgent: Device Recall" letter dated May 19, 2009 was issued to customers. Customers were advised to discontinue use of all affected E-cylinder holders. Consignees were asked to return the "Product Accountability Form" included with the recall letter, after which consignees will be provided with an improved cylinder holder at no cost. Direct questions to Stryker Communications at 1-972-410-7100.