Roche/Hitachi Modular E Module immunoassay analyzer, Roche Diagnostics, Ind., IN; GMMI Nos. 04998642001 and 03617505001.

This recall has been terminated (originally issued October 22, 2008).

Company
Roche Diagnostics Corp.
Recall Initiated
June 24, 2008
Posted
October 22, 2008
Terminated
October 22, 2009
Recall Number
Z-0165-2009
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

A software bug may result in pipetting from an incorrect reagent pack and/or assigning calibration curve parameters incorrectly.

Distribution

Nationwide Distribution --- including states of Florida, New Jersey and North Carolina.

Lot / Code Info

All units with software version 08-02.

Root Cause

Other

Action Taken

The consignees were notified of the problem via letter dated 6/24/08 and were provided instructions for a workaround.