Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2)

This recall has been terminated (originally issued May 9, 2006).

Company
General Electric Med Systems Llc
Recall Initiated
October 8, 2004
Posted
May 9, 2006
Terminated
December 31, 2006
Recall Number
Z-0854-06
Quantity
199
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

Display in fused mode- viewer allows 2 studies ( PET/CT) registered in the same spatial domain to be viewed simultaneously. These 2 images should be visually aligned to a precision of less than 1/2 PET Voxel. A problem associated with this version of software may introduce a shift of up to 1.5 PET Voxels, which can be as much as 8 mm with certain zoom factors.

Distribution

Distributed in 32 states: AK, AL, AZ, CA, CO, CT, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, NC, ND, NH, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV; Australia, Belgium, Brazil, Canada, Switzerland, Chile, Nigeria, China, France, Egypt, Ireland, Israel, Japan, Korea, the Netherlands, Norway, New Zealand, Phillipines, Sweden, and Taiwan.

Lot / Code Info

100943VP0, 67392II1, 36521VE3, 70827VE1, 419453BU6436332BU1, 70800VE8, 44586VE6, 341GE8, 97197VP8, 99813VP8, 97983VP1, 70820VE6, 58626VE3, 70795VE0, FRB3510J25, 58630VE5, 58624VE8, 58589VE3, 7823VE0, 44579VE1, 44597VE3, 44585VE8, 25125VE6, 425022BU1, 70796VE8, 93086VE7, 37621BU4, 76906VE7, 76899VE4, 70798VE4, 65504VE3, 65603VE3, 36543VE7, 58628VE9, 70811VE5, 44584VE1, 44581VE7, FRB3430VBM, 44589VE0, 44598VE1, 44582VE5, 44594VE0, 72740VE4, 1189GE0, 2030GE5, 2085GE0, 2025GE5, 72477II3, 72480II7, 343GE4, 97203VP4, 98272VP8, 423277BU3, 349GE1, 425018BU9, FRB3300M6G, 99699VP1, 98371VP8, 101216VP0, 425015BU5, 237GE8, 364GE0, 339GE2, 101196VP4, 419382BU7, 419383BU5, 366GE5, 367GE3, 357GE4, 2085GE9, 419457BU7, 25128VE0, 98436VP9, 99981VP3, 36513VE0, 98623UP2, 101000VP8, 2028GE9, 2799GE5, 1089GE2, 419450BU2, 101280VP6, 67269VE1, 100855VP6, 66250II2, 346GE7, 363GE2, 345GE9, 1073VP5, 101170VP9, 98516VP8, 2802GE7, 419356BU1, 66244II5, 1487GE8, 67389II7, 365GE7, 425021BU3, 383GE0, 25132VE2, 101334VP1, 95972VP6, 419449BU4, 2031GE3, 2800GE1, 36517VE1, FRB33605L9, 1024GE9, 95307VP5, 102235VP9, 102232VP6, 425204BU5, 1124GE7, 1122GE1, 419455BU1, 419456BU9, 102220VP1, 102555VP0, 97622VP5, 72249II6, 67210VE5, FRB33605KX, 419257BU1, 95268VP9, 338GE4, 101404VP2, 2090GE9, 2087GE5, 101308VP5, 101251VP7, 421989BU5, 342GE6, 99627VP2, 97933VP6, 98653VP9, 66249II4, 72479II9, 25126VE4, 356GE6, 25130VE6, 65450VE9, DE31500056, 347GE5, 421367BU4, 100912VP5, 25123VE1, 25124VE9, 99867VP4, DE24600031, 99910VP2, 101246VP7, 419357BU9, 101171VP7, 102825VP7, 100664VP2, 102856VP2, 424101BU4, 425020BU5, 354GE1, 36518VE9, 100885VP3, 2801GE9, USV41104ST, 2091GE7, USV41103MJ, 99994VP6, 102133VP6, 419549BU1, 25137VE1, 102078VP3, 425351BU4, 1066GE0, 98816VP2, 97591VP2, 99631VP4, 423318BU5, 2086GE7, 97905VP4, 100690VP7, 350GE9, 25139VE7, 1120GE5, 100001VP7, 2084GE2, 391228BU4, 101190VP7, 99909VP4, 1123GE9, 2803GE5, 98474VP0, 25127VE2, 67384II8, 98815VP4, 101288VP9, 101287VP1, 66241II1, 25134VE8

Root Cause

Other

Action Taken

An Urgent Safety Notice letter, dated October 8, 2004,(5121309-147, rev.E) was distributed to affected customers and a validated software patch that corrected the problem was installed on affected systems via Field Modification Instruction (FMI) 80114.