Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2)
This recall has been terminated (originally issued May 9, 2006).
- Company
- General Electric Med Systems Llc
- Recall Initiated
- October 8, 2004
- Posted
- May 9, 2006
- Terminated
- December 31, 2006
- Recall Number
- Z-0854-06
- Quantity
- 199
- Firm Location
- Waukesha, WI
- Official Source
- View on FDA website ↗
Reason for Recall
Display in fused mode- viewer allows 2 studies ( PET/CT) registered in the same spatial domain to be viewed simultaneously. These 2 images should be visually aligned to a precision of less than 1/2 PET Voxel. A problem associated with this version of software may introduce a shift of up to 1.5 PET Voxels, which can be as much as 8 mm with certain zoom factors.
Distribution
Distributed in 32 states: AK, AL, AZ, CA, CO, CT, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, NC, ND, NH, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV; Australia, Belgium, Brazil, Canada, Switzerland, Chile, Nigeria, China, France, Egypt, Ireland, Israel, Japan, Korea, the Netherlands, Norway, New Zealand, Phillipines, Sweden, and Taiwan.
Lot / Code Info
100943VP0, 67392II1, 36521VE3, 70827VE1, 419453BU6436332BU1, 70800VE8, 44586VE6, 341GE8, 97197VP8, 99813VP8, 97983VP1, 70820VE6, 58626VE3, 70795VE0, FRB3510J25, 58630VE5, 58624VE8, 58589VE3, 7823VE0, 44579VE1, 44597VE3, 44585VE8, 25125VE6, 425022BU1, 70796VE8, 93086VE7, 37621BU4, 76906VE7, 76899VE4, 70798VE4, 65504VE3, 65603VE3, 36543VE7, 58628VE9, 70811VE5, 44584VE1, 44581VE7, FRB3430VBM, 44589VE0, 44598VE1, 44582VE5, 44594VE0, 72740VE4, 1189GE0, 2030GE5, 2085GE0, 2025GE5, 72477II3, 72480II7, 343GE4, 97203VP4, 98272VP8, 423277BU3, 349GE1, 425018BU9, FRB3300M6G, 99699VP1, 98371VP8, 101216VP0, 425015BU5, 237GE8, 364GE0, 339GE2, 101196VP4, 419382BU7, 419383BU5, 366GE5, 367GE3, 357GE4, 2085GE9, 419457BU7, 25128VE0, 98436VP9, 99981VP3, 36513VE0, 98623UP2, 101000VP8, 2028GE9, 2799GE5, 1089GE2, 419450BU2, 101280VP6, 67269VE1, 100855VP6, 66250II2, 346GE7, 363GE2, 345GE9, 1073VP5, 101170VP9, 98516VP8, 2802GE7, 419356BU1, 66244II5, 1487GE8, 67389II7, 365GE7, 425021BU3, 383GE0, 25132VE2, 101334VP1, 95972VP6, 419449BU4, 2031GE3, 2800GE1, 36517VE1, FRB33605L9, 1024GE9, 95307VP5, 102235VP9, 102232VP6, 425204BU5, 1124GE7, 1122GE1, 419455BU1, 419456BU9, 102220VP1, 102555VP0, 97622VP5, 72249II6, 67210VE5, FRB33605KX, 419257BU1, 95268VP9, 338GE4, 101404VP2, 2090GE9, 2087GE5, 101308VP5, 101251VP7, 421989BU5, 342GE6, 99627VP2, 97933VP6, 98653VP9, 66249II4, 72479II9, 25126VE4, 356GE6, 25130VE6, 65450VE9, DE31500056, 347GE5, 421367BU4, 100912VP5, 25123VE1, 25124VE9, 99867VP4, DE24600031, 99910VP2, 101246VP7, 419357BU9, 101171VP7, 102825VP7, 100664VP2, 102856VP2, 424101BU4, 425020BU5, 354GE1, 36518VE9, 100885VP3, 2801GE9, USV41104ST, 2091GE7, USV41103MJ, 99994VP6, 102133VP6, 419549BU1, 25137VE1, 102078VP3, 425351BU4, 1066GE0, 98816VP2, 97591VP2, 99631VP4, 423318BU5, 2086GE7, 97905VP4, 100690VP7, 350GE9, 25139VE7, 1120GE5, 100001VP7, 2084GE2, 391228BU4, 101190VP7, 99909VP4, 1123GE9, 2803GE5, 98474VP0, 25127VE2, 67384II8, 98815VP4, 101288VP9, 101287VP1, 66241II1, 25134VE8
Root Cause
Other
Action Taken
An Urgent Safety Notice letter, dated October 8, 2004,(5121309-147, rev.E) was distributed to affected customers and a validated software patch that corrected the problem was installed on affected systems via Field Modification Instruction (FMI) 80114.