SJM Confirm, Implantable Cardiac Monitor (ICM) Models: DM2100, DM2102 (IDE). This product is a lead less non-therapeutic implantable cardiac monitor and is implanted to aid clinicians in diagnosing intermittent symptoms which may be due to arrhythmia

This recall has been terminated (originally issued July 5, 2011).

Company
St Jude Medical Crmd
Recall Initiated
March 31, 2011
Posted
July 5, 2011
Terminated
July 19, 2011
Recall Number
Z-2733-2011
Quantity
83 units
Firm Location
Sylmar, CA
Official Source
View on FDA website ↗

Reason for Recall

The recall was initiated because St. Jude Medical made new software available for the SJM Confirm DM2100 implantable cardiac monitor (ICM) that incorporates enhancements to the devices sensing algorithm. These enhancements are intended to provide improved sensitivity and specificity for diagnosing arrhythmias. The user is presented with the option to upgrade new software into the SJM Confirm ICM

Distribution

Worldwide distribution: (USA) Nationwide including the States of CA, WA, FL, WV, NY, MD and LA, and the country of Germany

Lot / Code Info

Confirm DM2100, DM2102 (IDE). Software Version 10.1.1.2

Root Cause

Software change control

Action Taken

St. Jude Cardiac Rhythm Management Division sent an "Important Product Correction" letter on March 25, 2011 to all affected customers. The letter explains the reason for the recall and provided recommendations including a software upgrade by SJM Field Clinical Engineer or Technical Services. Customers were recommended to fill out and return a "Physician Acknowledgement Form". For additional information please contact your local St.Jude Medical Sales Representative, Field Clinical Engineer or the Technical Services Department at 800-722-3774