RecallDepth

St. Jude Medical Riata ST Silicone Endocardial Defibrillation Leads Riata(7Fr), Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042 Product Usage: These Silicone transvenous high voltage (HV) leads connect an implantable cardioverter-defibrillator (lCD or CRT-D) to cardiac tissue to monitor and regulate a patients heart rate by providing pacing therapy and/or tachyarrhythmia therapy.

This recall has been terminated (originally issued January 11, 2012).

Company
St Jude Medical Crmd
Recall Initiated
November 28, 2011
Posted
January 11, 2012
Terminated
May 9, 2012
Recall Number
Z-0458-2012
Quantity
128,000 units total
Firm Location
Sylmar, CA
Official Source
View on FDA website ↗

Reason for Recall

The recall was initiated because St. Jude Medical has confirmed the failures associated with all cause insulation failure on the St. Jude Medical Riata and Riata ST Silicone Endocardial Defibrillation Leads with specific emphasis on externalized conductors.

Distribution

USA (Nationwide) and worldwide to: Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hungary, Iceland, Italy, Romania, Russia, Netherlands, Norway, Poland, Portugal, Russia, Serbia, Slovakia, Spain, Sweden, Switzerland, The Netherlands, UK, Australia, China, Hong Kong, Indonesia, Japan, Korea, Malaysia, Philippines, Singapore, Taiwan, Thailand, Vietnam, New Zealand, Argentina, Brazil, Colombia, Costa Rica, Ecuador, Mexico, Puerto Rico, Uruguay, Egypt, Greece, India, Iran, Israel, Joran, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, South Africa, Turkey, Malta, Cyprus, Iraq, Egypt, Lybia, Yemen, and U.A.E.

Lot / Code Info

Riata ST models (700X, 701X, 704X)

Root Cause

Device Design

Action Taken

St. Jude Medical sent a Medical Device Advisory Important Product Information Update letter dated November 28, 2011, Physician Communication dated December 15, 2010, Product Performance Report Extracts, and a Physician Device Advisory Notification to physicians by FedEx priority overnight service for receipt on November 29, 2011. The letters provides the customers with an explanation of the problem, the product, and the actions to be taken. The issuance of the recall letter is in conjunction with St. Jude's recent release November 2011 Product Performance Report (available online at http://sjmprofessional.com). The letter also explains about the Lead Peformance Summary, Root Cause, Clinical Implications, Rate of Occurrence from Complaints and Returns, New Peer Reviewed Publications, Data on Optim Insulated Leads (Durata and Riata ST Optim), and Recommendations and Mitigations. Physicians with any questions or concerns can contact their local St. Jude Medical representative or the Technical Services Department at (800) 722-3774.

More recalls by St Jude Medical Crmd

View all recalls by this company →