PKG, ALLIS DOUBLE ROW TEETH, P/N 0250080592 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
This recall has been terminated (originally issued December 12, 2014).
- Company
- Stryker Endoscopy
- Recall Initiated
- November 17, 2014
- Posted
- December 12, 2014
- Terminated
- July 17, 2015
- Recall Number
- Z-0717-2015
- Quantity
- 180,573 units total
- Firm Location
- San Jose, CA
- Official Source
- View on FDA website ↗
Reason for Recall
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lot / Code Info
140289 641421 840358 848483 940108 940131 946000 1040117 1040987 1041626 1042737 1042745 1045058 1046928 1143580 1145635 1145811 1145930 1146540 1241586 1242142 1242643 1243443 1243554 1244523 1340562 1341469 1341471 1341995 1343407 1343706 1344781 1346104 1346430 1346651 1346652
Root Cause
Error in labeling
Action Taken
Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.