RecallDepth

SERFAS 90 degree Energy Probe, Part Number 279-350-101; SERFAS Energy Probes are indicated for arthroscopic procedures of the knee, shoulder, ankle, hip, elbow and wrist. Specifically, the probes are used for resection, ablation and coagulation of soft tissue, as well as the hemostasis of blood vessels.

This recall has been terminated (originally issued June 24, 2015).

Company
Stryker Endoscopy
Recall Initiated
June 3, 2015
Posted
June 24, 2015
Terminated
February 10, 2016
Recall Number
Z-1831-2015
Quantity
22,063 devices
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

Stryker Endoscopy is recalling all non expired SERFAS 90 degree Energy Probes due to reports of fragments of the probe breaking off into the patient.

Distribution

Worldwide Distribution - US (nationwide) and Internationally to US Argentina, Australia, Bolivia, Brazil, Chile, China, Colombia, Hong Kong, India, Italy, Japan, Republic of Korea, Malaysia, Netherlands, Poland, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

Lot / Code Info

Part number 279-350-101; All non expired product; lot numbers 13128AE2 through 14337AE2.

Root Cause

Device Design

Action Taken

Stryker sent an Urgent Medical Device Recall letter dated June 5, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to do the following: 1. Inform individuals within their organization who need to be aware of this device removal. 2. Review inventory of lots of part number 279-350-101 and determine if you have the affected product (all non-expired devices) in stock. Response is required. 3. If no product is found, complete acknowledgement form located on the Stryker Endoscopy recall website endorecall.stryker.com by logging in using the account number and zip code on this letter. 4. If you do have product, segregate the product and call Stryker customer service at 1-800-624-4422 Option 3 to arrange for product return and issuance of credit. For questions regarding this recall call 408-754-2664.