PKG, BABCOCK FORCEPS, P/N 0250080321. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

This recall has been terminated (originally issued December 12, 2014).

Company
Stryker Endoscopy
Recall Initiated
November 17, 2014
Posted
December 12, 2014
Terminated
July 17, 2015
Recall Number
Z-0684-2015
Quantity
180,573 units total
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Lot / Code Info

140571 140572 140590 442506 542896 643523 646838 741305 741307 744923 842628 842632 842638 842640 847869 847870 847872 847873 847877 847878 847880 847881 847885 847886 1041276 1042935 1042951 1045190 1046166 1046900 1046901 1140664 1140821 1142174 1142665 1143446 1143728 1144692 1145109 1146164 1146165 1146388 1146900 1146901 1147163 1240407 1240955 1241953 1241954 1242409 1243807 1244132 1244241 1244441 1245480 1246060 1246405 1340290 1341254 1341835 1341836 1343535 1343993 1346493 1440592 1440788 1440789

Root Cause

Error in labeling

Action Taken

Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.