Amplilink (AL) Software 3.2.3 for use with COBAS AMPLICOR, CAP/CA, COBAS TaqMan,and COBAS AmpliPrep/COBAS TaqMan. The software can be used with a Laboratory Information System.

This recall has been terminated (originally issued September 18, 2012).

Company
Roche Molecular Systems, Inc.
Recall Initiated
April 23, 2012
Posted
September 18, 2012
Terminated
February 12, 2014
Recall Number
Z-2403-2012
Quantity
1292 instruments
Firm Location
Branchburg, NJ
Official Source
View on FDA website ↗

Reason for Recall

When using a Laboratory Information system (LIS) with automated systems utilizing AMPLILINK software v3.2 series (AL v3.2) there is a potential for wrong results to be assigned to an order.

Distribution

Nationwide (USA) Distribution

Lot / Code Info

05666678001

Root Cause

Device Design

Action Taken

Roche Molecular Systems Inc. sent "Urgent Field Safety Notice" letters and Faxback forms on 4/23/2012 via UPS to all affected customers. The issue was described and recommended actions were provided. A Product Advisory Notice and an Updated Field Safety Notice were issued as well.