Amplilink (AL) Software 3.2.3 for use with COBAS AMPLICOR, CAP/CA, COBAS TaqMan,and COBAS AmpliPrep/COBAS TaqMan. The software can be used with a Laboratory Information System.

Company
Roche Molecular Systems, Inc.
Recall Initiated
April 23, 2012
Posted
September 18, 2012
Terminated
February 12, 2014
Recall Number
Z-2403-2012
Quantity
1292 instruments
Firm Location
Branchburg, NJ

Reason for Recall

When using a Laboratory Information system (LIS) with automated systems utilizing AMPLILINK software v3.2 series (AL v3.2) there is a potential for wrong results to be assigned to an order.

Distribution

Nationwide (USA) Distribution

Lot / Code Info

05666678001

Root Cause

Device Design

Action Taken

Roche Molecular Systems Inc. sent "Urgent Field Safety Notice" letters and Faxback forms on 4/23/2012 via UPS to all affected customers. The issue was described and recommended actions were provided. A Product Advisory Notice and an Updated Field Safety Notice were issued as well.