Lumen Biomedical Xtract Aspiration Catheter system, intravascular catheter, consisting of One Catheter, Two 30 mL Syringes, One Extension Tube with Stopcock, and One 40 um Cell Strainer. Single-user, 0.014" guidewire compatible intravascular extraction and aspiration catheter, Part number 1450-0001 size 4.2F, and part number 1451-003 size 5.8F, distributed by Volcano Corporation.

Company
Volcano Corporation
Recall Initiated
November 3, 2010
Posted
November 23, 2010
Terminated
December 7, 2011
Recall Number
Z-0445-2011
Quantity
4724 units
Firm Location
Rancho Cordova, CA

Reason for Recall

Potential for product failure. Packaging/Storage issue results in compromised integrity and potential product failure during use, leading to potential patient injury.

Distribution

Worldwide distribution: USA, Greece, Egypt, Kuwait, Netherlands, Saudi Arabia, Italy, France, Sweden, Portugal, Great Britain, Austria, Switzerland, Germany, Romania, Poland, and Bulgaria

Lot / Code Info

Part number 1450-001 size 4.2F and 1451-003 size 5.8F, all lots.

Root Cause

Device Design

Action Taken

Firm initiated recall on 11/03/2010, and sent notifications to its consignees via letter. The firm has requested that consignees provide the notification to all personnel within the organization and letters will be sent directly to known customers on 11/04/2010 by Volcano. Customers have been instructed to cease use and contact customer service to arrange returns, and credit will be issued.