Baxter Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009

This recall is currently active, issued November 8, 2024. It was issued by Baxter Healthcare Corporation.

Company
Baxter Healthcare Corporation
Recall Initiated
October 2, 2024
Posted
November 8, 2024
Recall Number
Z-0348-2025
Quantity
8 units
Firm Location
Deerfield, IL
Official Source
View on FDA website ↗

Reason for Recall

The door on the Spectrum IQ Infusion pump may not be able to fully close.

Distribution

GA, ID

Lot / Code Info

00085412610900, Serial Numbers: 3796578, 3797382, 3797464, 3797578, 3797728, 3797733, 3798019, 3798200

Root Cause

Process control

Action Taken

Baxter Healthcare notified its consignees on 10/03 and 04/2024 via telephone. The call included information regarding the problem with the device, risk to health, and requested the affected units be removed from service pending inspection. On-site inspections to assess for impact and offer replacement, were scheduled.