RecallDepth

Diasol Acid Concentrate REF G100325-10Dex 100

This recall is currently active, issued October 23, 2024. It was issued by Diasol, Inc.

Company
Diasol, Inc
Recall Initiated
July 27, 2022
Posted
October 23, 2024
Recall Number
Z-0207-2025
Quantity
852 gallon bottles
Firm Location
Phoenix, AZ
Official Source
View on FDA website ↗

Reason for Recall

Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.

Distribution

U.S. Nationwide distribution in the state of AZ.

Lot / Code Info

Catalog Number: G100325-10Dex100 UDI code: B467100325101004 Lot Number:PHS06011/061722

Root Cause

Labeling mix-ups

Action Taken

On 07/26/2022, the firm communicated via email with customer that 2 case were incorrectly labeled. Cases marked 3K 2.5ca have inside box label as 3K 2.0ca. Customer was instructed not to use the product and the firm picked up mislabeled product.