RecallDepth

Brand Name: Diasol Product Name: 100230-10-DEX100, Liquid Concentrate for Bicarbonate Dialysis Non Pyrogenic, 2K, 3.0CA, 1.0MG, 208.2 liters

Company
Diasol, Inc
Recall Initiated
November 5, 2025
Posted
December 23, 2025
Recall Number
Z-0990-2026
Quantity
4,400 gallons
Firm Location
Phoenix, AZ

Reason for Recall

Safety and efficacy of dialysis acid concentrate cannot be assured

Distribution

US Nationwide distribution in the states of MO, CA, AZ, TX, IL, MA.

Lot / Code Info

UDI: None/Lot Code: PHV07222,PHV09082

Root Cause

Process design

Action Taken

On November 5, 2025, Diasol issued a "Urgent Medical Device Recall Notification to affected consignees. In addition to informing consignees about the recall, Diasol asked consignees to take the following actions: 1. Please return all product you have at your facility. 2. Please complete the attached acknowledgement form. 3. If product has been further distributed to other distributors, customer or sites, notify those consignees and provide them with a copy of the recall notification. 4. The notice should be provided to all staff members who use the product. 5. In the event you have not received adequate products from Rockwell you may continue to use Diasol Products at risk.

More recalls by Diasol, Inc

View all recalls by this company →