RecallDepth

Immunoassay Premium Controls (3x4x5mL and 12x5mL packages)

Company
Randox Laboratories
Recall Initiated
August 11, 2017
Terminated
March 29, 2019
Recall Number
Z-0969-2018
Quantity
117 Kits
Firm Location
Kearneysville, WV

Reason for Recall

The ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.

Distribution

Distribution US nationwide.

Lot / Code Info

Catalog numbers:  IA2633 (Lots: 398703, 402366) IA2638 (Lot 1666EC) IA2639 (Lots: 1486EC, 1516EC, 1626EC) IA2640 (Lots 1517EC, 1668EC)

Root Cause

Device Design

Action Taken

On July 24, 2017 an Urgent notice was issued instructing customers to place a copy a notice in all kits remaining in stock. This notifies customers that they should amend the claims for open stability in the instructions for use, to state use immediately after the 30 minute reconstitution period. Questions or concerns can be emailed to technical.services@randox.com

More recalls by Randox Laboratories

View all recalls by this company →