VERIGENE Nucleic Acid Test Kit-BC-GN Extraction Tray - 9.73 mL, REF 20-009-021
- Company
- Luminex Corporation
- Recall Initiated
- February 15, 2022
- Recall Number
- Z-1109-2022
- Quantity
- 984 units
- Firm Location
- Northbrook, IL
Reason for Recall
Higher than normal "No Call Internal Control failures" due to a missing reagent within a small subset of the extraction tray lot.
Distribution
US
Lot / Code Info
Lot Number 101321021B
Root Cause
Under Investigation by firm
Action Taken
Luminex sent a product recall notice on 02/15/2022 by email. The notice informed the user of "No Call Internal Control failures" and requested that they contact Luminex Global Support Services at support@luminexcorp.com to arrange for replacement and confirm the disposal of any kits in inventory.