Verigene Gram Positive Blood Culture (BC-GP) Nucleic Acid Test

Company
Luminex Corporation
Recall Initiated
June 20, 2024
Posted
August 8, 2024
Recall Number
Z-2533-2024
Quantity
1,100 kits
Firm Location
Northbrook, IL

Reason for Recall

Potential for defective test cartridges which may result in false positive results for the Staphylococcus lugdunensis and Enterococcus faecium targets.

Distribution

US Nationwide distribution including in the states of AZ, CA, CO, DE, FL, GA, IA, IL, KY, MD, MI, MN, MO, MS, NE, NV, OH, OK, SC, TN, TX, VA, VT, WA, WV.

Lot / Code Info

GTIN Number: 00840487101575; Catalog Number: 20-006-018; Lot Number: 031824018A; Expiration Date: 09/16/2024

Root Cause

Under Investigation by firm

Action Taken

Diasorin sent a recall notification on 06/20/2024, via email. Consignees were instructed to examine inventory and quarantine any affected units, arrange for replacement and confirm disposal of any of affected units on hand. Customers were asked to perform a retrospective review of results for VERIGENE Gram-Positive Blood Culture Nucleic Acid Test (BC-GP) to look for any false positives, distribute the notification to any relevant personnel, notify consignees if further distributed and complete and return the "Acknowledgement & Receipt Form."