RecallDepth

Stimuplex A, 30 DEG, 22GX2", 0.70x50mm, Catalogue Number: 4894502

Company
B Braun Medical Inc
Recall Initiated
May 17, 2024
Posted
June 11, 2024
Recall Number
Z-2134-2024
Quantity
1,157,640 units
Firm Location
Breinigsville, PA

Reason for Recall

The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Lot / Code Info

Catalogue Number: 4894502; UDI/DI: 04022495101310; Lot/Batch: 20E01, 20E03, 21E07, 21E09, 19E10, 22E10, 22E11, 21E13, 20E19, 20E20, 19E21, 19E22, 20E22, 23E26, 20E27, 23E31, 19F12, 19F13, 19F15, 19G03, 19G04, 19G06, 19G09, 19G20, 19G21, 19G24, 19G25, 19G27, 19H23, 19H28, 19H30, 19I11, 19I21, 19I25, 19I26, 19K11, 19K16, 19K17, 19K24, 19K25, 19L12, 19L13, 19L15, 19L19, 19M12, 19M14, 19M17, 19M21, 19M24, 19M26, 20A30, 20A31, 20B15, 20B18, 20B19, 20B21, 20B26, 20B27, 20C14, 20C17, 20C18, 20D01, 20D02, 20D04, 20D14, 20D16, 20D17, 20D21, 20D29, 20F06, 20F10, 20F11, 20F12, 20F13, 20G21, 20G22, 20G23, 20G29, 20G31, 20H01, 20H05, 20H26, 20H27, 20H29, 20L27, 20L28, 20M03, 20M19, 20M23, 20M25, 20M26, 21A16, 21A19, 21B04, 21B06, 21B10, 21B18, 21B20, 21B25, 21C11, 21C13, 21C17, 21D07, 21D08, 21D10, 21D29, 21F19, 21F23, 21F24, 21F25, 21F26, 21F30, 21G08, 21G09, 21G10, 21G14, 21G17, 21G20, 21G21, 21H05, 21H06, 21H07, 21H26, 21H28, 21I04, 21I22, 21I26, 21K07, 21K09, 21K13, 21K29, 21K30, 21L16, 21L17, 21L18, 21L19, 21L20, 21L30, 21M02, 21M03, 21M11, 21M14, 22A08, 22A25, 22B25, 22B27, 22C02, 22C03, 22C25, 22C29, 22D14, 22D15, 22D19, 22D21, 22F02, 22F04, 22G14, 22G16, 22G21, 22G23, 22G29, 22H25, 22I01, 22I03, 22I08, 22I28, 22K01, 22K05, 22K16, 22K18, 22K19, 22K20, 22L01, 22L02, 22L15, 22L16, 22M06, 22M15, 22M17, 23A08, 23A11, 23A31, 23B01, 23B07, 23B25, 23B28, 23C16, 23C18, 23C21, 23D19, 23D22, 23D28, 23F07, 23F30, 23G07, 23G11, 23G29, 23H01, 23I05, 23I06, 23I07, 23I23, 23I26, 23I28, 23I29, 23K18, 23K19.

Root Cause

Component design/selection

Action Taken

On May 17, 2024, URGENT MEDICAL DEVICE CORRECTION letters were sent to consignees. Actions Required by B. Braun Medical Inc. (BBMI) Customer/User: 1. Review this notice in its entirety. Ensure that all users in your organization of the above-mentioned products, and other concerned personnel are informed about this voluntary correction. Post this notification where the affected products are stored. a. If you are a distributor and have further distributed the product, please forward this notice to your consignees. The correction is to be extended to the consumer level. b. If you are a distributor that distributes partial cases, please ensure a copy of this notice is forwarded to your consignees. 2. Determine your current inventory. Review your inventory frequently to check these lot numbers. 3. Return the completed Urgent Medical Device Correction Acknowledgement Form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to recalls@bbraunusa.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). For questions regarding the label and impacted Stimuplex A Needles, please call Customer Support at 1-800-227-2862. Adverse reactions or quality problems experienced with this product, or questions about this correction may be reported to BBMI s Postmarket Surveillance Department by calling 1-833-425-1464. Adverse Reactions or quality problems in the United Sates may also be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

More recalls by B Braun Medical Inc

View all recalls by this company →